MagBead Cleanup (16S Only):

Some plates were cleaned up using the Nimbus platform protocol “AxyPrep MagBead PCR1 No MM”

Manually, it was done:

  • Equilibrate Beads to room Temperature

  • Add 24 ul of MagBeads to each well and 15 ul of replicate to same well of replicate

  • Pipette mix up and down 10 times.

  • Incubate at RT for 5 minutes

  • Secure plate on magnet plate; incubate at RT for 5 minutes (until wells are clear)

  • Remove 65 ul from each well; keep tips to left or right depending on the column to avoid bead pellet.

  • Add 100 ul Fresh 80% EtOH to each well. Incubate 30 seconds. Remove 100 ul from each well

  • Add 100 ul Fresh 80% EtOH to each well. Incubate 30 seconds. Remove 100 ul from each well

  • Reaspirate from each well to assure maximum EtOH removal

  • Allow plate to air dry for 7 minutes.

  • Remove sample plate from magnet plate.

  • Add 40 ul TE; pipette mix 10+ times. Incubate 2 minutes at RT.

  • Place sample plate back on magnet for 5 minutes or until all wells are cleared.

  • Transfer 40 ul to labeled transparent plate (Plate name_MIDs)

...