Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.
Info

02-04-2022 Illumina agreed there was likely a consumables problem. They are sending a replacement cartridge next week.

...

%Q30 Read 1

78.8

%Q30 Read 2

75.2

%Clusters PF

39

%Occupancy

90.2

%Clusters Passing Filter is concerning. We will reach out to Illumina.

...

  • 100/Results = ul of Pool to Add

100/10 4.98 = 10uL 20uL of Pool to Add

  • 100 - uL of Pool to Add = ul of “10 mM Tris 8.5” to Add

100- 10 20 = 90uL 80uL of 10mM Tris 8.5

Dilute 1 nM full pool to loading concentration of 50 pM:

...