Aliquot 11 ul into each well of an 8 pcr-tube strip. Add 4 ul IlluminaTest primers to 2. Add 4 ul 16STest primers to 1; add 2 ul 16STest primers to the next and 2 ul TE. Add 4 ul ITSTest primers to 1; add 2 ul ITSTest primers the next and 2 ul TE., Add 2 ul 16STest primers and 2 ul MC to the last 2. Add Test date to TE source.

Cap, Vortex, and spin down. Run on thermocycler program GSAF36:

...

  • Remove reagents and tape out of fridge for 30 minutes

  • Vortex and spin down Buffer and Ladder

  • Add 2 ul High Sensitivity Buffer to 9 wells of 2 empty Agilent Optical Tube Strips

  • Add 2 ul Ladder to the first well of one.

  • Add 2 ul sample to each well you added Buffer.

  • Apply caps, Label tops, and quick spin down.

  • Transport to INBRE.

  • Open TapeStation Software

  • Add Tape to slot with QR code in the back right

  • Vortex tube strips for 1 minute at 2000 rpm

    • Assign sample names in software

  • Spin tubes strips down for 1 minute

  • Add samples info instrument with ladder in A1 (back left)

  • After run completion, Create Report and add to report subpage

    • email or use usb to transfer data

  • After use, save any remaining MM in 2 ml tube. Transfer appropriate label for post sequencing check if necessary.