Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 2 Next »

Shannon Harris Muhammad Saqib

  • Vortex and quick spin 96 tubes and load into a 96-position tube rack.

  • Label Tube Rack 4Initials# (First Plate from Abby Hoffman: 4AH1

  • Use the 50 ml Integra expandable spacing pipette to transfer 30 ul from each tube into a transparent pcr plate. Label this plate the same as the tube rack.

  • Vortex and spin control oligo pool (16S and ITS coligo 0.01 pg/ul each AND 16S SG_GR and ITS SG_GR 0.03 pg/ul each. This should be in the left hand refrigerator. Send me a picture of the label to be certain. It should be the only plate in their marked 0.01 pg/ul coligo and 0.03 pg/ul synthgene.

  • Add 6 ul control oligo plate to the sample aliquot plate in a well to well fashion (A1 to A1, B1 to B1, . . .) This can be done using an 8 channel or the 96 channel. If you use the 8 channel, you can reuse the dome caps on the control plate as long as you avoid contamination. Return control plate.

  • Seal, vortex, and spin aliquot plate.

  • Check concentration on Synergy HTX and save file on petalibrary. Make a new folder under DNA Quantification labeled “NovaSeq4”

  • No labels