Keep MasterMix in the dark as much as possible. The qPCR dyes are light sensitive and degrade with exposure.

Add MasterMix and Samples:

Add 16 ul of Mastermix to each well needed.

Add 4 ul of each sample.

Setup Instrument For Run

  • Open “7500 System Software”

  • Hit New File Icon

  • Name Plate Appropriately (Copy from Confluence Page in Question). Hit “Next”.

  • Select “Unknown”. Hit “Add”. Click “Next”

  • Highlight All Wells To Be Used And Assign “Unknown” Detector

  • Assign All Wells Their Task From The Dropdown next to “Unknown” detector. Choices Are “NTC”, “Unknown”, or “Standard”. Click “Finish”

  • Assign Standards Values:

    • Std1: 20, Std2: 2, Std3: 0.2, Std4: 0.02, Std5: 0.002, Std6: 0.0002

  • Navigate to “Instrument” tab and change cycle parameters to:

...