qPCR Checks for Illumina Library Preps

qPCR Checks for Illumina Library Preps

Dilute your samples 1 to 1000:

The KAPA Library Quant Kit (Illumina) that we use has a limited range of detectability. Too high leads to signal saturation and too low leads to not enough signal. The standards range from 20 pM to 0.0002 pM. Our library prep methods produce products in the 1-100 nM range. Generally, all samples must be diluted 1 to 1000.

Add 1 ul sample to 999 ul ultra pure water. Vortex and quick centrifuge sample dilutions.

Create MasterMix:

Generally, all samples, standards, and ntcs should be done in triplicate. Enter the appropriate number into the top “# of rxns” cell below.

Keep MasterMix in the dark as much as possible. The qPCR dyes are light sensitive and degrade with exposure.

Add MasterMix and Samples:

Add 16 ul of Mastermix to each well needed.

Add 4 ul of each sample.

Setup Instrument For Run

  • Open “7500 System Software”

  • Hit New File Icon

  • Name Plate Appropriately (Copy from Confluence Page in Question). Hit “Next”.

  • Select “Unknown”. Hit “Add”. Click “Next”

  • Highlight All Wells To Be Used And Assign “Unknown” Detector

  • Assign All Wells Their Task From The Dropdown next to “Unknown” detector. Choices Are “NTC”, “Unknown”, or “Standard”. Click “Finish”

  • Assign Standards Values:

    • Std1: 20, Std2: 2, Std3: 0.2, Std4: 0.02, Std5: 0.002, Std6: 0.0002

  • Navigate to “Instrument” tab and change cycle parameters to:

Temp C

Cycles

Time (m:ss)

Temp C

Cycles

Time (m:ss)

94

1X

0:01

95

1X

5:00

95

35X

0:30*

60

35X

1:00*

*These might change if we create any significantly longer products.

  • Hit the Save icon and ensure that it is being saved as an “.sds” document in the folder “C:/Applied Biosystems/SDS Documents”

  • Double Check Layout and Instrument Tabs Are Correct.

  • Press circular indent in the darker gray region to eject tray. Load Plate with A1 lined up with A1 mark.

  • Press “Start” in software. If Start is not an option, close “7500 System Software” and reopen. Reopen the file you just created and try “Start” again.

  • Always turn instrument off when finished for the day.

Related content

1 -step Library preparation protocol for 16s or ITS
1 -step Library preparation protocol for 16s or ITS
Read with this
qPCR of pools and phiX
qPCR of pools and phiX
More like this
DNA quantification using the Biotek Synergy HTX
DNA quantification using the Biotek Synergy HTX
Read with this
GBS MasterMix Prep and QC protocol
GBS MasterMix Prep and QC protocol
More like this
2 -step Library preparation protocol for 16s or ITS
2 -step Library preparation protocol for 16s or ITS
Read with this
Amplicon MasterMix Prep and QC protocol
Amplicon MasterMix Prep and QC protocol
More like this