The library pool was flagged by Psomagen as suspect. They could not get it to qPCR and the size was too small.
Enlarged Electropherogram from Psomagen
We initially tried qPCR on the full strength pool and a 1:1000 dilution.
ul/rxn | Reagent | # of rxns | ul needed |
---|---|---|---|
10 ul | KAPA SYBR FAST qPCR MM (2X) | 24 | 300 |
2 ul | Primer Premix (10X) | 24 | 60 |
4 ul | Ultra Pure Water | 24 | 120 |
16 ul | Total Volume | 24 | 480 |
Add 4 ul of templates
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | |
---|---|---|---|---|---|---|---|---|---|---|---|---|
A | NTC | NTC | NTC | 1:1000 N2 | ||||||||
B | 0.0002 pM Std | 0.0002 pM Std | 0.0002 pM Std | 1:1000 N2 | ||||||||
C | 0.002 pM Std | 0.002 pM Std | 0.002 pM Std | 1:1000 N2 | ||||||||
D | 0.02 pM Std | 0.02 pM Std | 0.02 pM Std | |||||||||
E | 0.2 pM Std | 0.2 pM Std | 0.2 pM Std | |||||||||
F | 2 pM Std | 2 pM Std | 2 pM Std | |||||||||
G | 20 pM Std | 20 pM Std | 20 pM Std | |||||||||
H | NovaSeq2 | NovaSeq2 | NovaSeq2 |
N2=NovaSeq2
The results confirmed the low concentration reported. We did get amplification though.
The full strength gave a concentration of 6 pMoles/ul and the dilution gave a thousandth of that confirming each other.
We tried to concentrate the library, but after researching the requirements for the NovaSeq loading concentration realized that would not work.
We went back to some of the individual plates to investigate possible problems. We redid PCR2 of one of the first plates done and one of the last from the saved and frozen PCR1s and compared them with the original PCR1s via qPCR of 1:1000 dilutions.
ul/rxn | Reagent | # of rxns | ul needed |
---|---|---|---|
10 ul | KAPA SYBR FAST qPCR MM (2X) | 53 | 600 |
2 ul | Primer Premix (10X) | 53 | 120 |
4 ul | Ultra Pure Water | 53 | 240 |
16 ul | Total Volume | 53 | 960 |
Add 4 ul template:
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | |
---|---|---|---|---|---|---|---|---|---|---|---|---|
A | NTC | NTC | NTC | PE1 A1 | JC1 A1 | PE1 A1 Take 2 | JC1 A1 Take2 | |||||
B | 0.0002 pM Std | 0.0002 pM Std | 0.0002 pM Std | PE1 B1 | JC1 B1 | PE1 B1 Take 2 | JC1 B1 Take 2 | |||||
C | 0.002 pM Std | 0.002 pM Std | 0.002 pM Std | PE1 C1 | JC1 C1 | PE1 C1 Take 2 | JC1 C1 Take 2 | |||||
D | 0.02 pM Std | 0.02 pM Std | 0.02 pM Std | PE1 D1 | JC1 D1 | PE1 D1 Take 2 | JC1 D1 Take 2 | |||||
E | 0.2 pM Std | 0.2 pM Std | 0.2 pM Std | PE1 E1 | JC1 E1 | PE1 E1 Take 2 | JC1 E1 Take 2 | |||||
F | 2 pM Std | 2 pM Std | 2 pM Std | PE1 F1 | JC1 F1 | PE1 F1 Take 2 | JC1 F1 Take 2 | |||||
G | 20 pM Std | 20 pM Std | 20 pM Std | PE1 G1 | JC1 G1 | PE1 G1 Take 2 | JC1 G1 Take 2 | |||||
H | PE1 H1 | JC1 H1 | PE1 H1 Take 2 | JC1 H1 Take 2 |
Well, there is certainly a difference between the initial PCR2 and the redone version. There is a 100 fold increase in the redone version.
We will repeat this one more time with the flow cell primers created anew from the 300 uM stock using one plate, PE1. We will also do a repeated PCR1 for this plate to include in the qPCR comparison.