SOP for Dionex Ion Chromatography HPIC System
SOP for Dionex Ion Chromatography HPIC System Version 1.2
Approved by laboratory manager on 1-31-2022
For Running the Anion side
Log on to the Anion Computer
Username: Ion Chromatography
Password: EcologyBGC
Click on Chromeleon Software
Click on instrument tab
Select “Integrion_Anions” on the left (if it isn’t selected automatically)
(tabs at the top, left to right)
Sampler *do not touch*
Click on “Pump_ECD” and do the following:
Flow rate – 1.0mL/minute
Type in gray boxes and hit the “enter” key
Turn On Eluent Generator
30.00mM concentration
Turn On General On
Turn on CR_TC Anions
Turn on Suppressor Anions
Column Temperature should be 30
When CD total is less than 1microS you can run
Everything should have a green light
Click “Take Control” at the top bar
If you cannot take control then someone else forgot to “Release Control”
Data Tab On lower left
Look for the folders “Integrion_Anions” on the Anion Computer and expand them using the small “+” sign
Locate your name and date
If you do not have a folder yet, right click on the “Integrion_Anions” folder and create a folder with your name and date
Then locate the “Templates as of 2022” folder and expand it
Open the “Anion Sequences” folder using the “+” symbol
Copy over the sequence named “Anions Full Tray Run with calibration 11-3-2021” in the “Anion Sequence” folder and paste to your name folder that is under “Integrion_Anions” and rename the sequence to include the date of the run first then your name and whatever you want to name your sequence.
Example: Jake, needs to run 120 samples for his pharmaceuticals class. They copy the sequence name “Anions Full Tray Run with calibration 11-3-2021”, and put it under the folder “Jake 2022”.
They rename the sequence to “1-12-2022 Jake Pharmaceuticals Tray 1”To setup your schedule
The templates should start with a calibration, if you do not need the calibrations, you can delete them.
If you only need a partial tray delete, the number of samples you don’t need.
Your adjusted templates should start with a “blank” and a “CCV” or if you need calibrations, it will start with “Cal Std 1”
Following the “Blank” and “CCV” are 10 samples, numbered from 1 through 10 before the “CCV” and “Blank” need to be run again.
Be sure to name your samples with your initials followed by a number/letter sequence, that would replace the place holder 1 through 10.
For example, Jake Adams would be “JA01” instead of “1”.
At the end of your samples run a CCV A followed by a Blank
Double check to make sure you assigned a position to every sample to include the CCV and Blank(s)
Click “Save” at the top of the schedule
Loading Samples
Push the carousel release button and wait for the arm to come up above the carousel tray.
The carousel tray can now be moved freely and locate the position number that the sample will match up with.
Once everything is loaded and the schedule is completed and double checked for accuracy, go back to the “Instruments Tab” and then to the “Queue” tab of the Anion instrument.
Add your Anion sequence to the instrument by clicking “Add” and navigating to the “Instrument Data” folder and then to the “Integrion Anions” folder, then locate the folder with your name and select the correct sequence
Click “Ready Check” on the right side
If no errors, and the Ultra Purified Water containers on each instrument are full and the waste container is ¾ or less full, proceed to hit “Start” on the right side to start running your samples.
Click “Release Control” at the top bar if you are completed with your samples on the Anion instrument.
For running the Cation side
Log on to the Cation Computer
Username: Ion Chromatography
Password: EcologyBGC
Click on Chromeleon Software
Click on instrument tab
Select “Integrion_Cations” on the left (if it isn’t selected automatically)
(tabs at the top, left to right)
Sampler *do not touch*
Pump for Cations
Flow rate – 0.64mL/minute
Turn On Eluent Generator
35.00mM concentration
Turn On General On
Turn on CR_TC Cations
Turn on Suppressor Cations
When CD total is less than 1microS you can run
Column Temperature should be 40
Everything should have a green light
Click “Take Control” at the top bar
If you cannot take control then someone else forgot to “Release Control”
Data Tab On lower left
Look for the folders “Integrion_Cations” on the Cations Computer and expand them using the small “+” sign
Locate your name and date
If you do not have a folder yet, right click on the “Integrion_Cations”, folder and create a folder with your name and date
Then locate the “Templates as of 2022” folder and expand it
Open the “Cation Sequences” folder using the + symbol
Copy over the sequence named “Cations-Full ASAP CCV Blank Template 11-3-2021” in the “Cation Sequence” folder to your name folder that is under “Integrion_Cation” and rename the sequence to include the date of the run first then your name and whatever you want to name your sequence.
Example: Jake, needs to run 120 samples for his pharmaceuticals class. They copy the sequence name “Cations-Full ASAP CCV Blank Template 11-3-2021”, and put it under the folder “Jake 2022”.
They rename the sequence to “1-12-2022 Jake Pharmaceuticals Tray 1”To setup your schedule
The templates should start with a calibration, if you do not need the calibrations, you can delete them.
If you only need a partial tray delete the number of samples you don’t need.
Your adjusted templates should start with a “blank” and a “CCV” or if you need calibrations, it will start with “Cal Std 1”
Following the “Blank” and “CCV” are 10 samples, numbered from 1 through 10 before the “CCV” and “Blank” need to be run again.
Be sure to name your samples with your initials followed by a number/letter sequence, that would replace the place holder 1 through 10.
For example, Jake Adams would be “JA01” instead of “1”.
At the end of your samples run a CCV A followed by a Blank
Double check to make sure you assigned a position to every sample to include the CCV and Blank(s)
Click “Save” at the top of the schedule
Loading Samples
Push the carousel release button and wait for the arm to come up above the carousel tray.
The carousel tray can now be moved freely and locate the position number that the sample will match up with.
Once everything is loaded and the schedule is completed and double checked for accuracy, go back to the “Instruments Tab” and then to the “Queue” tab of the Cation instrument.
Add your Cation sequence to the instrument by clicking “Add” and navigating to the “Instrument Data” folder and then to the “Integrion Cation” folder, then locate the folder with your name and select the correct sequence
Click “Ready Check” on the right side
If no errors, and the Ultra Purified Water containers on each instrument are full and the waste container is ¾ or less full, proceed to hit “Start” on the right side to start running your samples.
Click “Release Control” at the top bar if you are completed with your samples on the Cation instrument.