Fragment Analysis

Fragment Analysis

  • The Agilent 2100 Bioanalyzer is available for biological fragment analysis.

  • DNA 1000 Chips are the only diagnostic chips kept on hand currently.

    • Detection range 15 to 1500 bp

    • Sample quantity range 0.5 - 50 ng/ul

  • Samples should be diluted to within the working range of 1-50 ng/ul

  • An aliquot of your suspension buffer should be included with your submission

  • Your expected band(s) should fall within the 15-1500 bp range

  • 10 ul per sample is the minimum acceptable volume for submissions

  • Protocol for Fragment Analysis on Agilent 2100 Bioanalyzer

 