LIMS Testing Notes
@Nicholas Case
GTL functionality notes
Print laminated sheets with barcoded common suffixes so we can just scan the additions instead of typing
Check Out
Having to click the “Add” button after every scan is not desirable
DNA Extraction (Plate)
Add some Extraction Kit options
DNeasy 96 Blood & Tissue Kit
I wasn’t seeing this previously, but now after entering a suffix for the new plate in “Plate Barcode” it gives me a warning that prevents me from continuing, “?Positions unmapped”. When I click on “View container” it appears I have to drag each well over in order to proceed. That will not work.
DNA Quantification
These files appear to be correct. However, when I use them to upload into the plate reader software, something bizarre happens with the first entry. It is changed to this: Gregg0001. I do not know where those extra characters are coming from, but I imagine they are not allowing the resultant file to pass muster for the next step.
DNA Normalization
I think this is just because the previous step didn’t go right, but the “Retrieve and Parse File” button for GreggPl01_Stock did not retrieve a file. This problem is persisting. Though, I have a suspected culprit.
The suspected culprit, , is the criminal. I proceeded with GreggPlate03 erasing those characters and this step produced the correct file.
PCR Library Prep
This seems to be good
MagBead Cleanup
I cannot get this to mark step as complete when I set the 2 16S or ITS plates as going into the same new barcoded plate. It just seems to stall and never mark step as complete.
This appears to be because somehow I got my earlier attempt to not error when entering the same barcode for the two plates combined into one. This time it tells me that the plate already exists. I am still stuck here
h
General
Add “Aliquot Plate” as a Protocol
Create daughter plate from any plate to carry through next steps
My CR2600 (earlier model adds a pre and post \ to each scan. Need to find out how to stop that and make it press enter.